Abstract

Powered lower-limb prostheses have the potential to assist amputees in the push-off phase during level walking, as well as stair and slope ascent. Compared to electromechanical actuation, the advantages of using eletrohydrostatic actuation (EHA) including high power density, low noise and good controllability. Especially for the application in lower limb prosthetic joints, an EHA provides a quick and smooth switch between passive and active operation modes. This paper presents the testing results using a new eletrohydrostatic powered ankle prosthesis which combines an EHA with a controllable passive damper. The new powered ankle prosthesis has been tested by both a transtibial and a transfemoral amputee. The test results show that the ankle prosthesis can provide sufficient power to assist toe push-off and subsequent foot lift (dorsiflexion) and operating passively in the rest of a gait cycle. The test results are compared with the ankle dorsiflexion angle and torque for healthy subjects. The timing control method and the performance of the EHA are discussed.
Original languageEnglish
Pages (from-to)185-191
Number of pages7
JournalIFAC-PapersOnLine
Volume49
Issue number21
DOIs
Publication statusE-pub ahead of print - 10 Nov 2016

Keywords

  • EHA
  • powered ankle prosthesis

Cite this

Testing an Electrohydrostatic Powered Ankle Prosthesis with Transtibial and Transfemoral Amputees. / Yu, Tian; Plummer, Andrew; Iravani, Pejman; Bhatti, Jawaad; Zahedi, Saeed; Moser, David.

In: IFAC-PapersOnLine, Vol. 49, No. 21, 10.11.2016, p. 185-191.

Research output: Contribution to journalArticle

Yu, Tian ; Plummer, Andrew ; Iravani, Pejman ; Bhatti, Jawaad ; Zahedi, Saeed ; Moser, David. / Testing an Electrohydrostatic Powered Ankle Prosthesis with Transtibial and Transfemoral Amputees. In: IFAC-PapersOnLine. 2016 ; Vol. 49, No. 21. pp. 185-191.
@article{a437092882854c29b01870bb2ddbede8,
title = "Testing an Electrohydrostatic Powered Ankle Prosthesis with Transtibial and Transfemoral Amputees",
abstract = "Powered lower-limb prostheses have the potential to assist amputees in the push-off phase during level walking, as well as stair and slope ascent. Compared to electromechanical actuation, the advantages of using eletrohydrostatic actuation (EHA) including high power density, low noise and good controllability. Especially for the application in lower limb prosthetic joints, an EHA provides a quick and smooth switch between passive and active operation modes. This paper presents the testing results using a new eletrohydrostatic powered ankle prosthesis which combines an EHA with a controllable passive damper. The new powered ankle prosthesis has been tested by both a transtibial and a transfemoral amputee. The test results show that the ankle prosthesis can provide sufficient power to assist toe push-off and subsequent foot lift (dorsiflexion) and operating passively in the rest of a gait cycle. The test results are compared with the ankle dorsiflexion angle and torque for healthy subjects. The timing control method and the performance of the EHA are discussed.",
keywords = "EHA, powered ankle prosthesis",
author = "Tian Yu and Andrew Plummer and Pejman Iravani and Jawaad Bhatti and Saeed Zahedi and David Moser",
year = "2016",
month = "11",
day = "10",
doi = "10.1016/j.ifacol.2016.10.543",
language = "English",
volume = "49",
pages = "185--191",
journal = "IFAC-PapersOnLine",
issn = "1474-6670",
publisher = "Elsevier",
number = "21",

}

TY - JOUR

T1 - Testing an Electrohydrostatic Powered Ankle Prosthesis with Transtibial and Transfemoral Amputees

AU - Yu, Tian

AU - Plummer, Andrew

AU - Iravani, Pejman

AU - Bhatti, Jawaad

AU - Zahedi, Saeed

AU - Moser, David

PY - 2016/11/10

Y1 - 2016/11/10

N2 - Powered lower-limb prostheses have the potential to assist amputees in the push-off phase during level walking, as well as stair and slope ascent. Compared to electromechanical actuation, the advantages of using eletrohydrostatic actuation (EHA) including high power density, low noise and good controllability. Especially for the application in lower limb prosthetic joints, an EHA provides a quick and smooth switch between passive and active operation modes. This paper presents the testing results using a new eletrohydrostatic powered ankle prosthesis which combines an EHA with a controllable passive damper. The new powered ankle prosthesis has been tested by both a transtibial and a transfemoral amputee. The test results show that the ankle prosthesis can provide sufficient power to assist toe push-off and subsequent foot lift (dorsiflexion) and operating passively in the rest of a gait cycle. The test results are compared with the ankle dorsiflexion angle and torque for healthy subjects. The timing control method and the performance of the EHA are discussed.

AB - Powered lower-limb prostheses have the potential to assist amputees in the push-off phase during level walking, as well as stair and slope ascent. Compared to electromechanical actuation, the advantages of using eletrohydrostatic actuation (EHA) including high power density, low noise and good controllability. Especially for the application in lower limb prosthetic joints, an EHA provides a quick and smooth switch between passive and active operation modes. This paper presents the testing results using a new eletrohydrostatic powered ankle prosthesis which combines an EHA with a controllable passive damper. The new powered ankle prosthesis has been tested by both a transtibial and a transfemoral amputee. The test results show that the ankle prosthesis can provide sufficient power to assist toe push-off and subsequent foot lift (dorsiflexion) and operating passively in the rest of a gait cycle. The test results are compared with the ankle dorsiflexion angle and torque for healthy subjects. The timing control method and the performance of the EHA are discussed.

KW - EHA

KW - powered ankle prosthesis

U2 - 10.1016/j.ifacol.2016.10.543

DO - 10.1016/j.ifacol.2016.10.543

M3 - Article

VL - 49

SP - 185

EP - 191

JO - IFAC-PapersOnLine

JF - IFAC-PapersOnLine

SN - 1474-6670

IS - 21

ER -