Ten-year in vivo wear measurement of a fully congruent mobile bearing unicompartmental knee arthroplasty

A. J. Price, A. Short, C. Kellett, D. Beard, H. S. Gill, H. Pandit, C. A. Dodd, D. W. Murray

Research output: Contribution to journalArticle

66 Citations (Scopus)

Abstract

Polyethylene particulate wear debris continues to be implicated in the aetiology of aseptic loosening following knee arthroplasty. The Oxford unicompartmental knee arthroplasty employs a spherical femoral component and a fully congruous meniscal bearing to increase contact area and theoretically reduce the potential for polyethylene wear. This study measures the in vivo ten-year linear wear of the device, using a roentgenstereophotogrammetric technique. In this in vivo study, seven medial Oxford unicompartmental prostheses, which had been implanted ten years previously were studied. Stereo pairs of radiographs were acquired for each patient and the films were analysed using a roentgen stereophotogrammetric analysis calibration and a computer-aided design model silhouette-fitting technique. Penetration of the femoral component into the original volume of the bearing was our estimate of linear wear. In addition, eight control patients were examined less than three weeks post-insertion of an Oxford prosthesis, where no wear would be expected. The control group showed no measured wear and suggested a system accuracy of 0.1 mm. At ten years, the mean linear wear rate was 0.02 mm/year. The results from this in vivo study confirm that the device has low ten-year linear wear in clinical practice. This may offer the device a survival advantage in the long term.
Original languageEnglish
Pages (from-to)1493-1497
Number of pages5
JournalJournal of Bone and Joint Surgery - British Volume
Volume87
Issue number11
DOIs
Publication statusPublished - 2005

Fingerprint

Knee Replacement Arthroplasties
Polyethylene
Thigh
Equipment and Supplies
Prostheses and Implants
Computer-Aided Design
Calibration
Control Groups
Survival

Cite this

Ten-year in vivo wear measurement of a fully congruent mobile bearing unicompartmental knee arthroplasty. / Price, A. J.; Short, A.; Kellett, C.; Beard, D.; Gill, H. S.; Pandit, H.; Dodd, C. A.; Murray, D. W.

In: Journal of Bone and Joint Surgery - British Volume, Vol. 87, No. 11, 2005, p. 1493-1497.

Research output: Contribution to journalArticle

Price, A. J. ; Short, A. ; Kellett, C. ; Beard, D. ; Gill, H. S. ; Pandit, H. ; Dodd, C. A. ; Murray, D. W. / Ten-year in vivo wear measurement of a fully congruent mobile bearing unicompartmental knee arthroplasty. In: Journal of Bone and Joint Surgery - British Volume. 2005 ; Vol. 87, No. 11. pp. 1493-1497.
@article{c4a0326973b444ed97e1996b03534150,
title = "Ten-year in vivo wear measurement of a fully congruent mobile bearing unicompartmental knee arthroplasty",
abstract = "Polyethylene particulate wear debris continues to be implicated in the aetiology of aseptic loosening following knee arthroplasty. The Oxford unicompartmental knee arthroplasty employs a spherical femoral component and a fully congruous meniscal bearing to increase contact area and theoretically reduce the potential for polyethylene wear. This study measures the in vivo ten-year linear wear of the device, using a roentgenstereophotogrammetric technique. In this in vivo study, seven medial Oxford unicompartmental prostheses, which had been implanted ten years previously were studied. Stereo pairs of radiographs were acquired for each patient and the films were analysed using a roentgen stereophotogrammetric analysis calibration and a computer-aided design model silhouette-fitting technique. Penetration of the femoral component into the original volume of the bearing was our estimate of linear wear. In addition, eight control patients were examined less than three weeks post-insertion of an Oxford prosthesis, where no wear would be expected. The control group showed no measured wear and suggested a system accuracy of 0.1 mm. At ten years, the mean linear wear rate was 0.02 mm/year. The results from this in vivo study confirm that the device has low ten-year linear wear in clinical practice. This may offer the device a survival advantage in the long term.",
author = "Price, {A. J.} and A. Short and C. Kellett and D. Beard and Gill, {H. S.} and H. Pandit and Dodd, {C. A.} and Murray, {D. W.}",
year = "2005",
doi = "10.1302/0301-620X.87B11.16325",
language = "English",
volume = "87",
pages = "1493--1497",
journal = "Journal of Bone and Joint Surgery - British Volume",
issn = "0301-620X",
publisher = "British Editorial Society of Bone and Joint Surgery",
number = "11",

}

TY - JOUR

T1 - Ten-year in vivo wear measurement of a fully congruent mobile bearing unicompartmental knee arthroplasty

AU - Price, A. J.

AU - Short, A.

AU - Kellett, C.

AU - Beard, D.

AU - Gill, H. S.

AU - Pandit, H.

AU - Dodd, C. A.

AU - Murray, D. W.

PY - 2005

Y1 - 2005

N2 - Polyethylene particulate wear debris continues to be implicated in the aetiology of aseptic loosening following knee arthroplasty. The Oxford unicompartmental knee arthroplasty employs a spherical femoral component and a fully congruous meniscal bearing to increase contact area and theoretically reduce the potential for polyethylene wear. This study measures the in vivo ten-year linear wear of the device, using a roentgenstereophotogrammetric technique. In this in vivo study, seven medial Oxford unicompartmental prostheses, which had been implanted ten years previously were studied. Stereo pairs of radiographs were acquired for each patient and the films were analysed using a roentgen stereophotogrammetric analysis calibration and a computer-aided design model silhouette-fitting technique. Penetration of the femoral component into the original volume of the bearing was our estimate of linear wear. In addition, eight control patients were examined less than three weeks post-insertion of an Oxford prosthesis, where no wear would be expected. The control group showed no measured wear and suggested a system accuracy of 0.1 mm. At ten years, the mean linear wear rate was 0.02 mm/year. The results from this in vivo study confirm that the device has low ten-year linear wear in clinical practice. This may offer the device a survival advantage in the long term.

AB - Polyethylene particulate wear debris continues to be implicated in the aetiology of aseptic loosening following knee arthroplasty. The Oxford unicompartmental knee arthroplasty employs a spherical femoral component and a fully congruous meniscal bearing to increase contact area and theoretically reduce the potential for polyethylene wear. This study measures the in vivo ten-year linear wear of the device, using a roentgenstereophotogrammetric technique. In this in vivo study, seven medial Oxford unicompartmental prostheses, which had been implanted ten years previously were studied. Stereo pairs of radiographs were acquired for each patient and the films were analysed using a roentgen stereophotogrammetric analysis calibration and a computer-aided design model silhouette-fitting technique. Penetration of the femoral component into the original volume of the bearing was our estimate of linear wear. In addition, eight control patients were examined less than three weeks post-insertion of an Oxford prosthesis, where no wear would be expected. The control group showed no measured wear and suggested a system accuracy of 0.1 mm. At ten years, the mean linear wear rate was 0.02 mm/year. The results from this in vivo study confirm that the device has low ten-year linear wear in clinical practice. This may offer the device a survival advantage in the long term.

UR - http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&dopt=Citation&list_uids=16260665

UR - http://dx.doi.org/10.1302/0301-620X.87B11.16325

U2 - 10.1302/0301-620X.87B11.16325

DO - 10.1302/0301-620X.87B11.16325

M3 - Article

VL - 87

SP - 1493

EP - 1497

JO - Journal of Bone and Joint Surgery - British Volume

JF - Journal of Bone and Joint Surgery - British Volume

SN - 0301-620X

IS - 11

ER -