Polyethylene wear in Oxford unicompartmental knee replacement: a retrieval study of 47 bearings

B. J. Kendrick, D. Longino, H. Pandit, U. Svard, H. S. Gill, C. A. Dodd, D. W. Murray, A. J. Price

Research output: Contribution to journalArticle

50 Citations (Scopus)

Abstract

The Oxford Unicompartmental Knee replacement (UKR) was introduced as a design to reduce polyethylene wear. There has been one previous retrieval study involving this implant, which reported very low rates of wear in some specimens but abnormal patterns of wear in others. There has been no further investigation of these abnormal patterns. The bearings were retrieved from 47 patients who had received a medial Oxford UKR for anteromedial osteoarthritis of the knee. None had been studied previously. The mean time to revision was 8.4 years (sd 4.1), with 20 having been implanted for over ten years. The macroscopic pattern of polyethylene wear and the linear penetration were recorded for each bearing. The mean rate of linear penetration was 0.07 mm/year. The patterns of wear fell into three categories, each with a different rate of linear penetration; 1) no abnormal macroscopic wear and a normal articular surface, n = 16 (linear penetration rate = 0.01 mm/year); 2) abnormal macroscopic wear and normal articular surfaces with extra-articular impingement, n = 16 (linear penetration rate = 0.05 mm/year); 3) abnormal macroscopic wear and abnormal articular surfaces with intra-articular impingement +/- signs of non-congruous articulation, n = 15 (linear penetration rate = 0.12 mm/year). The differences in linear penetration rate were statistically significant (p <0.001). These results show that very low rates of polyethylene wear are possible if the device functions normally. However, if the bearing displays suboptimal function (extra-articular, intra-articular impingement or incongruous articulation) the rates of wear increase significantly.
Original languageEnglish
Pages (from-to)367-373
Number of pages7
JournalJournal of Bone and Joint Surgery - British Volume
Volume92
Issue number3
DOIs
Publication statusPublished - 2010

Fingerprint

Knee Replacement Arthroplasties
Polyethylene
Joints
Knee Osteoarthritis
Equipment and Supplies

Cite this

Polyethylene wear in Oxford unicompartmental knee replacement: a retrieval study of 47 bearings. / Kendrick, B. J.; Longino, D.; Pandit, H.; Svard, U.; Gill, H. S.; Dodd, C. A.; Murray, D. W.; Price, A. J.

In: Journal of Bone and Joint Surgery - British Volume, Vol. 92, No. 3, 2010, p. 367-373.

Research output: Contribution to journalArticle

Kendrick, B. J. ; Longino, D. ; Pandit, H. ; Svard, U. ; Gill, H. S. ; Dodd, C. A. ; Murray, D. W. ; Price, A. J. / Polyethylene wear in Oxford unicompartmental knee replacement: a retrieval study of 47 bearings. In: Journal of Bone and Joint Surgery - British Volume. 2010 ; Vol. 92, No. 3. pp. 367-373.
@article{bac10a33ee804408a7d3cc9753656556,
title = "Polyethylene wear in Oxford unicompartmental knee replacement: a retrieval study of 47 bearings",
abstract = "The Oxford Unicompartmental Knee replacement (UKR) was introduced as a design to reduce polyethylene wear. There has been one previous retrieval study involving this implant, which reported very low rates of wear in some specimens but abnormal patterns of wear in others. There has been no further investigation of these abnormal patterns. The bearings were retrieved from 47 patients who had received a medial Oxford UKR for anteromedial osteoarthritis of the knee. None had been studied previously. The mean time to revision was 8.4 years (sd 4.1), with 20 having been implanted for over ten years. The macroscopic pattern of polyethylene wear and the linear penetration were recorded for each bearing. The mean rate of linear penetration was 0.07 mm/year. The patterns of wear fell into three categories, each with a different rate of linear penetration; 1) no abnormal macroscopic wear and a normal articular surface, n = 16 (linear penetration rate = 0.01 mm/year); 2) abnormal macroscopic wear and normal articular surfaces with extra-articular impingement, n = 16 (linear penetration rate = 0.05 mm/year); 3) abnormal macroscopic wear and abnormal articular surfaces with intra-articular impingement +/- signs of non-congruous articulation, n = 15 (linear penetration rate = 0.12 mm/year). The differences in linear penetration rate were statistically significant (p <0.001). These results show that very low rates of polyethylene wear are possible if the device functions normally. However, if the bearing displays suboptimal function (extra-articular, intra-articular impingement or incongruous articulation) the rates of wear increase significantly.",
author = "Kendrick, {B. J.} and D. Longino and H. Pandit and U. Svard and Gill, {H. S.} and Dodd, {C. A.} and Murray, {D. W.} and Price, {A. J.}",
year = "2010",
doi = "10.1302/0301-620X.92B3.22491",
language = "English",
volume = "92",
pages = "367--373",
journal = "Journal of Bone and Joint Surgery - British Volume",
issn = "0301-620X",
publisher = "British Editorial Society of Bone and Joint Surgery",
number = "3",

}

TY - JOUR

T1 - Polyethylene wear in Oxford unicompartmental knee replacement: a retrieval study of 47 bearings

AU - Kendrick, B. J.

AU - Longino, D.

AU - Pandit, H.

AU - Svard, U.

AU - Gill, H. S.

AU - Dodd, C. A.

AU - Murray, D. W.

AU - Price, A. J.

PY - 2010

Y1 - 2010

N2 - The Oxford Unicompartmental Knee replacement (UKR) was introduced as a design to reduce polyethylene wear. There has been one previous retrieval study involving this implant, which reported very low rates of wear in some specimens but abnormal patterns of wear in others. There has been no further investigation of these abnormal patterns. The bearings were retrieved from 47 patients who had received a medial Oxford UKR for anteromedial osteoarthritis of the knee. None had been studied previously. The mean time to revision was 8.4 years (sd 4.1), with 20 having been implanted for over ten years. The macroscopic pattern of polyethylene wear and the linear penetration were recorded for each bearing. The mean rate of linear penetration was 0.07 mm/year. The patterns of wear fell into three categories, each with a different rate of linear penetration; 1) no abnormal macroscopic wear and a normal articular surface, n = 16 (linear penetration rate = 0.01 mm/year); 2) abnormal macroscopic wear and normal articular surfaces with extra-articular impingement, n = 16 (linear penetration rate = 0.05 mm/year); 3) abnormal macroscopic wear and abnormal articular surfaces with intra-articular impingement +/- signs of non-congruous articulation, n = 15 (linear penetration rate = 0.12 mm/year). The differences in linear penetration rate were statistically significant (p <0.001). These results show that very low rates of polyethylene wear are possible if the device functions normally. However, if the bearing displays suboptimal function (extra-articular, intra-articular impingement or incongruous articulation) the rates of wear increase significantly.

AB - The Oxford Unicompartmental Knee replacement (UKR) was introduced as a design to reduce polyethylene wear. There has been one previous retrieval study involving this implant, which reported very low rates of wear in some specimens but abnormal patterns of wear in others. There has been no further investigation of these abnormal patterns. The bearings were retrieved from 47 patients who had received a medial Oxford UKR for anteromedial osteoarthritis of the knee. None had been studied previously. The mean time to revision was 8.4 years (sd 4.1), with 20 having been implanted for over ten years. The macroscopic pattern of polyethylene wear and the linear penetration were recorded for each bearing. The mean rate of linear penetration was 0.07 mm/year. The patterns of wear fell into three categories, each with a different rate of linear penetration; 1) no abnormal macroscopic wear and a normal articular surface, n = 16 (linear penetration rate = 0.01 mm/year); 2) abnormal macroscopic wear and normal articular surfaces with extra-articular impingement, n = 16 (linear penetration rate = 0.05 mm/year); 3) abnormal macroscopic wear and abnormal articular surfaces with intra-articular impingement +/- signs of non-congruous articulation, n = 15 (linear penetration rate = 0.12 mm/year). The differences in linear penetration rate were statistically significant (p <0.001). These results show that very low rates of polyethylene wear are possible if the device functions normally. However, if the bearing displays suboptimal function (extra-articular, intra-articular impingement or incongruous articulation) the rates of wear increase significantly.

UR - http://www.scopus.com/inward/record.url?scp=77649200553&partnerID=8YFLogxK

UR - http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&dopt=Citation&list_uids=20190307

UR - http://dx.doi.org/10.1302/0301-620X.92B3.22491

U2 - 10.1302/0301-620X.92B3.22491

DO - 10.1302/0301-620X.92B3.22491

M3 - Article

VL - 92

SP - 367

EP - 373

JO - Journal of Bone and Joint Surgery - British Volume

JF - Journal of Bone and Joint Surgery - British Volume

SN - 0301-620X

IS - 3

ER -