Abstract
Two new prototype assembly methods have been evaluated for biosensors that combine an integrated circuit (IC) sensor with a culture chamber. The first method uses a poly-ethylene glycol (PEG) mould to mask the IC sensor during application of a room temperature vulcanising (RTV) silicone elastomer used to insulate the bondpads and bondwires. The second method utilises the 'partial encapsulation' service offered by Quik-Pak, USA. Both methods were shown to provide good electrical insulation and demonstrated biocompatibility with the NG108-15 cell line. These methods are particularly useful for the assembly of low-cost ICs with a small (4 mm2) sensor area.
Original language | English |
---|---|
Pages (from-to) | 973-979 |
Number of pages | 7 |
Journal | Medical Engineering & Physics |
Volume | 33 |
Issue number | 8 |
DOIs | |
Publication status | Published - Oct 2011 |
Keywords
- packaging
- assembly
- partial encapsulation
- biosensor
- prototype
- integrated circuit