In vivo measurement of total knee replacement wear

C. F. Kellett, A. Short, A. Price, H. S. Gill, D. W. Murray

Research output: Contribution to journalArticlepeer-review

33 Citations (SciVal)

Abstract

Polyethylene wear is one of the most important causes of failure of total knee replacements (TKRs). Currently, wear can only be accurately measured by retrieval studies. There is a need for a method to measure wear accurately in vivo. We have developed a Roentgen stereophotogrammetric analysis (RSA) system that can measure penetration of the metallic femoral component into the polyethylene of the tibia. We have used this system to study six AGC TKRs at 6 years postoperatively and six control AGC TKRs at 2 weeks postoperatively. The mean difference between the RSA measured bearing thickness and the manufacturer's quoted values for the control group was −0.03 mm (S.D. 0.17). The average linear penetration in the study group was 0.8 mm (S.D. 0.46). This was significantly (P<0.0001) different from the control group. The average linear penetration rate was 0.13 mm per year (S.D. 0.08). We would expect the penetration to deepen with time. In young active patients, this could be a cause for concern, particularly with a thin bearing. The current system is accurate enough to measure wear at 5 years post TKR. It has the potential for predicting long-term wear problems with new designs of TKR and new materials within 2 years.
Original languageEnglish
Pages (from-to)183-187
Number of pages5
JournalThe Knee
Volume11
Issue number3
DOIs
Publication statusPublished - 2004

Fingerprint

Dive into the research topics of 'In vivo measurement of total knee replacement wear'. Together they form a unique fingerprint.

Cite this