Polyethylene wear is one of the most important causes of failure of total knee replacements (TKRs). Currently, wear can only be accurately measured by retrieval studies. There is a need for a method to measure wear accurately in vivo. We have developed a Roentgen stereophotogrammetric analysis (RSA) system that can measure penetration of the metallic femoral component into the polyethylene of the tibia. We have used this system to study six AGC TKRs at 6 years postoperatively and six control AGC TKRs at 2 weeks postoperatively. The mean difference between the RSA measured bearing thickness and the manufacturer's quoted values for the control group was −0.03 mm (S.D. 0.17). The average linear penetration in the study group was 0.8 mm (S.D. 0.46). This was significantly (P<0.0001) different from the control group. The average linear penetration rate was 0.13 mm per year (S.D. 0.08). We would expect the penetration to deepen with time. In young active patients, this could be a cause for concern, particularly with a thin bearing. The current system is accurate enough to measure wear at 5 years post TKR. It has the potential for predicting long-term wear problems with new designs of TKR and new materials within 2 years.
Original languageEnglish
Pages (from-to)183-187
Number of pages5
JournalThe Knee
Volume11
Issue number3
DOIs
Publication statusPublished - 2004

Fingerprint

Knee Replacement Arthroplasties
Polyethylene
Control Groups
Thigh
Tibia

Cite this

Kellett, C. F., Short, A., Price, A., Gill, H. S., & Murray, D. W. (2004). In vivo measurement of total knee replacement wear. The Knee, 11(3), 183-187. https://doi.org/10.1016/S0968-0160(03)00070-X

In vivo measurement of total knee replacement wear. / Kellett, C. F.; Short, A.; Price, A.; Gill, H. S.; Murray, D. W.

In: The Knee, Vol. 11, No. 3, 2004, p. 183-187.

Research output: Contribution to journalArticle

Kellett, CF, Short, A, Price, A, Gill, HS & Murray, DW 2004, 'In vivo measurement of total knee replacement wear', The Knee, vol. 11, no. 3, pp. 183-187. https://doi.org/10.1016/S0968-0160(03)00070-X
Kellett, C. F. ; Short, A. ; Price, A. ; Gill, H. S. ; Murray, D. W. / In vivo measurement of total knee replacement wear. In: The Knee. 2004 ; Vol. 11, No. 3. pp. 183-187.
@article{8e96a0f7475c45a0a54d4c39198dd50b,
title = "In vivo measurement of total knee replacement wear",
abstract = "Polyethylene wear is one of the most important causes of failure of total knee replacements (TKRs). Currently, wear can only be accurately measured by retrieval studies. There is a need for a method to measure wear accurately in vivo. We have developed a Roentgen stereophotogrammetric analysis (RSA) system that can measure penetration of the metallic femoral component into the polyethylene of the tibia. We have used this system to study six AGC TKRs at 6 years postoperatively and six control AGC TKRs at 2 weeks postoperatively. The mean difference between the RSA measured bearing thickness and the manufacturer's quoted values for the control group was −0.03 mm (S.D. 0.17). The average linear penetration in the study group was 0.8 mm (S.D. 0.46). This was significantly (P<0.0001) different from the control group. The average linear penetration rate was 0.13 mm per year (S.D. 0.08). We would expect the penetration to deepen with time. In young active patients, this could be a cause for concern, particularly with a thin bearing. The current system is accurate enough to measure wear at 5 years post TKR. It has the potential for predicting long-term wear problems with new designs of TKR and new materials within 2 years.",
author = "Kellett, {C. F.} and A. Short and A. Price and Gill, {H. S.} and Murray, {D. W.}",
year = "2004",
doi = "10.1016/S0968-0160(03)00070-X",
language = "English",
volume = "11",
pages = "183--187",
journal = "The Knee",
issn = "0968-0160",
publisher = "Elsevier",
number = "3",

}

TY - JOUR

T1 - In vivo measurement of total knee replacement wear

AU - Kellett, C. F.

AU - Short, A.

AU - Price, A.

AU - Gill, H. S.

AU - Murray, D. W.

PY - 2004

Y1 - 2004

N2 - Polyethylene wear is one of the most important causes of failure of total knee replacements (TKRs). Currently, wear can only be accurately measured by retrieval studies. There is a need for a method to measure wear accurately in vivo. We have developed a Roentgen stereophotogrammetric analysis (RSA) system that can measure penetration of the metallic femoral component into the polyethylene of the tibia. We have used this system to study six AGC TKRs at 6 years postoperatively and six control AGC TKRs at 2 weeks postoperatively. The mean difference between the RSA measured bearing thickness and the manufacturer's quoted values for the control group was −0.03 mm (S.D. 0.17). The average linear penetration in the study group was 0.8 mm (S.D. 0.46). This was significantly (P<0.0001) different from the control group. The average linear penetration rate was 0.13 mm per year (S.D. 0.08). We would expect the penetration to deepen with time. In young active patients, this could be a cause for concern, particularly with a thin bearing. The current system is accurate enough to measure wear at 5 years post TKR. It has the potential for predicting long-term wear problems with new designs of TKR and new materials within 2 years.

AB - Polyethylene wear is one of the most important causes of failure of total knee replacements (TKRs). Currently, wear can only be accurately measured by retrieval studies. There is a need for a method to measure wear accurately in vivo. We have developed a Roentgen stereophotogrammetric analysis (RSA) system that can measure penetration of the metallic femoral component into the polyethylene of the tibia. We have used this system to study six AGC TKRs at 6 years postoperatively and six control AGC TKRs at 2 weeks postoperatively. The mean difference between the RSA measured bearing thickness and the manufacturer's quoted values for the control group was −0.03 mm (S.D. 0.17). The average linear penetration in the study group was 0.8 mm (S.D. 0.46). This was significantly (P<0.0001) different from the control group. The average linear penetration rate was 0.13 mm per year (S.D. 0.08). We would expect the penetration to deepen with time. In young active patients, this could be a cause for concern, particularly with a thin bearing. The current system is accurate enough to measure wear at 5 years post TKR. It has the potential for predicting long-term wear problems with new designs of TKR and new materials within 2 years.

UR - http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&dopt=Citation&list_uids=15194093

UR - http://dx.doi.org/10.1016/S0968-0160(03)00070-X

U2 - 10.1016/S0968-0160(03)00070-X

DO - 10.1016/S0968-0160(03)00070-X

M3 - Article

VL - 11

SP - 183

EP - 187

JO - The Knee

JF - The Knee

SN - 0968-0160

IS - 3

ER -