Impedimetric approach for quantifying low bacteria concentrations based on the changes produced in the electrode-solution interface during the pre-attachment stage

X Munoz-Berbel, N Vigues, A T A Jenkins, J Mas, F J Munoz

Research output: Contribution to journalArticlepeer-review

46 Citations (SciVal)

Abstract

This paper describes an approach for quantifying low concentrations of bacteria, particularly Escherichia coli, based on the measurement of the initial attachment of bacteria to platinum surfaces, using impedance spectroscopy. The value of the interface capacitance in the pre-attachment stage (before 1 min of attachment) showed correlation with suspended concentration of bacteria from 10(1) to 10(7) CFU mL(-1) (colony fort-ning units per mL). This method was found to be sensitive to the attachment time, to the applied potential and to the size of the counter electrode. The sensor lifetime was also evaluated. (C) 2008 Elsevier B.V. All rights reserved.
Original languageEnglish
Pages (from-to)1540-1546
Number of pages7
JournalBiosensors and Bioelectronics
Volume23
Issue number10
DOIs
Publication statusPublished - 2008

Bibliographical note

ID number: ISI:000255793200018

Fingerprint

Dive into the research topics of 'Impedimetric approach for quantifying low bacteria concentrations based on the changes produced in the electrode-solution interface during the pre-attachment stage'. Together they form a unique fingerprint.