Abstract
Left ventricular assist devices (LVADs) have long been used to treat adults with heart failure, but LVAD options for pediatric patients with heart failure are lacking. Despite the urgent need for long-term, implantable pediatric LVADs, design
challenges such as hemolysis, pump thrombosis, and bleeding persist. We have developed a Hemocompatibility Assessment Platform (HAP) to identify blood trauma from individual LVAD components. A HAP would aid in refining pump components before in vivo testing, thereby preventing unnecessary animal
sacrifice and reducing development time and cost. So that the HAP does not confound hemolysis data, the HAP drive system consists of an enlarged air-gap motor coupled to a magnetic levitation system. Although it is known that an enlarged air gap motor will have diminished performance, while the larger gap in
the motor will cause less blood damage, the trade-offs are not fully characterized. Therefore, in this study we evaluated these trade-offs to determine an optimal rotor diameter for the HAP drive motor. The motor performance was characterized with an experimental method by determining the torque constant for the HAP drive motor with varied rotor diameters. The torque threshold was set as 10 mNm to achieve a nominal current of 3.5A. Hemolysis in the HAP drive motor gap was estimated by calculating scalar shear stress generated in the HAP motor gap
analytically and numerically. A design criterion of 30 Pa was selected for scalar shear stress to achieve minimal hemolysis and platelet activation in the HAP drive system.
challenges such as hemolysis, pump thrombosis, and bleeding persist. We have developed a Hemocompatibility Assessment Platform (HAP) to identify blood trauma from individual LVAD components. A HAP would aid in refining pump components before in vivo testing, thereby preventing unnecessary animal
sacrifice and reducing development time and cost. So that the HAP does not confound hemolysis data, the HAP drive system consists of an enlarged air-gap motor coupled to a magnetic levitation system. Although it is known that an enlarged air gap motor will have diminished performance, while the larger gap in
the motor will cause less blood damage, the trade-offs are not fully characterized. Therefore, in this study we evaluated these trade-offs to determine an optimal rotor diameter for the HAP drive motor. The motor performance was characterized with an experimental method by determining the torque constant for the HAP drive motor with varied rotor diameters. The torque threshold was set as 10 mNm to achieve a nominal current of 3.5A. Hemolysis in the HAP drive motor gap was estimated by calculating scalar shear stress generated in the HAP motor gap
analytically and numerically. A design criterion of 30 Pa was selected for scalar shear stress to achieve minimal hemolysis and platelet activation in the HAP drive system.