Objectives: Total artificial hearts (TAHs) are used for treatment of end-stage biventricular heart failure where a suitable donor heart is unavailable. Haemolysis is the damage to red blood cells with potentially harmful side effects. Mechanical circulatory support devices, like a TAH, create flow conditions that may give rise to haemolysis. Most haemolysis models are developed for rotary blood pumps. The aim of this study was to assess the appropriateness of available haemolysis models on a positive-displacement TAH.
Methods: A CFD model of the Realheart TAH V11.2 was created in Ansys Fluent V2022R2, using the same previously developed fluid-structure interaction modelling strategy that predicted valve motion for the legacy device V11C. The haemolysis index (HI, %) was calculated by assessing a shear-dependent power-law model using both an Eulerian scalar transport and Lagrangian particle tracking approach, using empirical constants established for rotary blood pumps. The model was solved until outlet HI values were constant. Three conditions were simulated: a baseline systemic condition at 5 L/min was compared against a systemic flow at 7 L/min and pulmonary flow at 5 L/min, achieved by varying heart rate and downstream two-element Windkessel parameters. Haemolysis was experimentally determined according to ASTM standard F1841-97.
Results: Pulmonary HI was 29% and 16% smaller than the baseline systemic HI for both Eulerian and Lagrangian models respectively. Systemic HI at 7 L/min was 24% and 30% greater than baseline for both Eulerian and Lagrangian models respectively. Experimentally, a 29% decrease and 80% increase between baseline, pulmonary and systemic flow at 7 L/min respectively was observed.
Conclusions: Eulerian and Lagrangian HI was the same order of magnitude. The relative change in HI between pulmonary and systemic baseline was the same as experiments, but the model did not exactly capture the same increase at 7 L/min. Future work will focus on model optimisation and parameter investigation.
Original languageEnglish
Publication statusPublished - 27 Aug 2023
EventEuropean Society for Artificial Organs (ESAO) 49th Annual Congress - Bergamo, Italy
Duration: 29 Aug 2023 → 1 Sept 2023

Conference

ConferenceEuropean Society for Artificial Organs (ESAO) 49th Annual Congress
Country/TerritoryItaly
CityBergamo
Period29/08/23 → 1/09/23

Fingerprint

Dive into the research topics of 'HAEMOLYSIS MODELLING OF A POSITIVE-DISPLACEMENT TOTAL ARTIFICIAL HEART'. Together they form a unique fingerprint.

Cite this