Preeclampsia affects 2 to 8% of pregnant women worldwide and accounts for 22% of the maternal mortality rate in the African region. Ultrasound shear wave elastography (US-SWE) has been used in vivo to detect placenta stiffness apparent in preeclampsia, despite questions regarding its accuracy. Utilizing placenta tissue-mimicking materials (TMMs) for ultrasound elastography can significantly advance placental research, integrating cellular and molecular biology to model gestational diseases. Placenta TMMs that replicate the heterogeneity, and acoustic and mechanical properties of the placenta but are also US-SWE compatible, are crucial for understanding factors that influence changing placental stiffness, the influence of gestational age, and quantifying its accuracy for this application in the clinic.
Preliminary research into these US-SWE TMMs investigated agar-based phantoms of varying concentrations to simulate placental tissue; however, low agar concentrations (2%) yielded Young’s modulus values nearly twice that of diseased placenta. Further reducing agar concentration affected phantom viability, while enhancements to acoustic properties compromised the mechanical integrity of these phantoms. This challenge highlighted the need to explore alternative TMMs for US-SWE.
This study investigated silk and Polyvinyl alcohol cryogel (PVA-c) as potential substitutes to replicate normal and diseased placental conditions associated with preeclampsia. Homogeneous silk and PVA-c phantoms were created and characterized for acoustic (attenuation and sound speed) and mechanical properties (stiffness). Measurements were taken from five locations within the phantoms using a Verasonics Vantage 256 ultrasound system with multiple-track shear wave elastography coupled with a 192-element linear transducer. An Instron 5965 facilitated mechanical testing to corroborate the Young's modulus estimation of stiffness.
The findings revealed a strong correlation between ultrasound-derived and mechanically measured stiffness and the phantoms’ acoustic properties, affirming their validity as realistic placenta models. Future research will delve into their long-term stability and cost-effectiveness with disease-specific features for understanding diseases through advanced imaging techniques.
Original languageEnglish
Publication statusPublished - 7 Jul 2025
EventLoke Centre for Trophoblast Research Annual Meeting 2025: The Placenta at Term - Churchill College, University of Cambridge, Cambridge, UK United Kingdom
Duration: 7 Jul 2025 → 8 Jul 2025
https://www.trophoblast.cam.ac.uk/files/loke_ctr_prospectus_2025_1.pdf

Conference

ConferenceLoke Centre for Trophoblast Research Annual Meeting 2025
Country/TerritoryUK United Kingdom
CityCambridge
Period7/07/25 → 8/07/25
Internet address

Funding

Funders
Commonwealth Scholarships Commission
Schlumberger Foundation, Faculty for the Future Fellowship Program

    UN SDGs

    This output contributes to the following UN Sustainable Development Goals (SDGs)

    1. SDG 3 - Good Health and Well-being
      SDG 3 Good Health and Well-being

    Fingerprint

    Dive into the research topics of 'Exploring silk and PVA-c as placenta tissue mimicking materials in ultrasound shear wave elastography for preeclampsia prediction'. Together they form a unique fingerprint.

    Cite this