Abstract
Objective: Implantable sacral anterior root stimulators enable bladder emptying after spinal cord injury but do not prevent reflex incontinence. A closed-loop
neuroprosthesis that detects and inhibits reflex bladder contractions could address this, but first, reliable detection of bladder fullness from the sacral roots.
Further, the distribution of afferent bladder activity between sacral roots, and the relationship between efferent and afferent activity within each root, remains
unclear and must be clarified to guide implant design.
Methods: Electrode books were implanted on the S1–S3 extra-dural sacral roots bilaterally in six terminally anesthetized sheep. Afferent electroneurogram
(ENG) was recorded concurrently from all implanted roots during filling cystometries and correlated with bladder pressure. Each root was individually electrically stimulated and the bladder pressure response recorded. Postmortem morphometric analysis determined fiber size distribution in each root.
Results: Overall, S2 ENG activity showed the highest correlation with bladder pressure, and electrical stimulation of S2 and S3 produced the greatest increases
in bladder pressure. Fiber size distribution did not correlate with either ENG activity or bladder pressure response. Significant variation was identified between
individual sheep, but notably, in four of six sheep, a single sacral root had both the highest ENG correlation to bladder pressure and the greatest bladder response to stimulation.
Significance: This study demonstrates reliable recording of bladder afferents from sacral roots using clinically applicable electrodes. It provides the first
systematic recording of bladder ENG concurrently across three pairs of sacral roots in multiple animals, and the first characterization of signal distribution
between roots. Significant individual variation is identified, impacting the design of future implantable sacral neuroprostheses for bladder control.
neuroprosthesis that detects and inhibits reflex bladder contractions could address this, but first, reliable detection of bladder fullness from the sacral roots.
Further, the distribution of afferent bladder activity between sacral roots, and the relationship between efferent and afferent activity within each root, remains
unclear and must be clarified to guide implant design.
Methods: Electrode books were implanted on the S1–S3 extra-dural sacral roots bilaterally in six terminally anesthetized sheep. Afferent electroneurogram
(ENG) was recorded concurrently from all implanted roots during filling cystometries and correlated with bladder pressure. Each root was individually electrically stimulated and the bladder pressure response recorded. Postmortem morphometric analysis determined fiber size distribution in each root.
Results: Overall, S2 ENG activity showed the highest correlation with bladder pressure, and electrical stimulation of S2 and S3 produced the greatest increases
in bladder pressure. Fiber size distribution did not correlate with either ENG activity or bladder pressure response. Significant variation was identified between
individual sheep, but notably, in four of six sheep, a single sacral root had both the highest ENG correlation to bladder pressure and the greatest bladder response to stimulation.
Significance: This study demonstrates reliable recording of bladder afferents from sacral roots using clinically applicable electrodes. It provides the first
systematic recording of bladder ENG concurrently across three pairs of sacral roots in multiple animals, and the first characterization of signal distribution
between roots. Significant individual variation is identified, impacting the design of future implantable sacral neuroprostheses for bladder control.
| Original language | English |
|---|---|
| Article number | 1857570 |
| Number of pages | 16 |
| Journal | Frontiers in Neuroscience |
| Volume | 20 |
| Early online date | 12 Jun 2026 |
| DOIs | |
| Publication status | Published - 12 Jun 2026 |
Acknowledgements
The authors would like to thank all staff in the Biological Services Unit, Royal Veterinary College, UK, for their assistance with the animal study. The authors would also like to thank Dr Piercy and the Comparative Neuromuscular Diseases Laboratory at the Royal Veterinary College for technical advice on morphometry and use of their laboratory facilities.Funding
The author(s) declared that financial support was received for this work and/or its publication. This work was funded by the Engineering and Physical Sciences Research Council through grant EP/P018947/1. MP received funding for this work through the Dr Brian Nicholson scholarship.
| Funders | Funder number |
|---|---|
| Engineering and Physical Sciences Research Council | EP/P018947/1 |
Keywords
- bladder afferent ENG
- Brindley device
- neuroprosthesis
- sacral anterior root stimulation
- sacral roots
- urinary bladder pressure
Fingerprint
Dive into the research topics of 'Distribution of bladder afferent activity across the sacral roots in sheep shows marked individual variation: Implications for neuroprosthesis design'. Together they form a unique fingerprint.Cite this
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS