Continuous glucose monitor overestimates glycemia, with the magnitude of bias varying by postprandial test and individual – A randomized crossover trial.

Katie Hutchins, James Betts, Dylan Thompson, Aaron Hengist, Javier Gonzalez

Research output: Contribution to journalArticlepeer-review

Abstract

Background: Continuous glucose monitors (CGM) are used to characterize postprandial glycemia, yet no study has directly tested how different test foods/beverages alter CGM accuracy.

Objective: Assess glycemic responses to test foods/drinks using CGM versus capillary sampling (criterion).

Methods: Fifteen healthy females(n=9) and males(n=6) completed 7 laboratory visits in a randomized crossover design with ≥48h washout between visits. During each visit, participants consumed an oral carbohydrate challenge comprising either 50g glucose(CONTROL), or equivalent carbohydrate as whole-fruits(WHOLE), blended-fruits(BLEND), commercial fruit smoothie(PRODUCT), commercial smoothie ingested over 30±4min(SLOW), commercial smoothie with 5g inulin(FIBER), commercial smoothie providing 30g carbohydrate(DOSE). Glycemia was recorded from both CGM and capillary samples every 15min for 120min and expressed as incremental areas under the curve(iAUC). Glycemic index(GI) was calculated relative to CONTROL where appropriate. Exploratory analyses examined 1) inter-individual heterogeneity of CGM bias versus criterion; and 2) whether CGM bias could be improved with adjustment for baseline differences.

Results: CGM-estimated fasting and postprandial glucose concentrations were(mean±SD) 0.9±0.6 and 0.9±0.5 mmol/L higher than capillary estimates, respectively(both, p<0.001). CGM bias varied by postprandial test such that GI for PRODUCT was higher with CGM(69, 95%CI: 48, 99) versus capillary(53, 95%CI: 40,69; P=0.05). Furthermore, differences in CGM versus capillary fasting glucose concentrations varied by participant(p=0.001). Unadjusted, CGM overestimated time >7.8mmol/L by ~4-fold, and adjustment for baseline differences reduced this overestimate to ~2-fold(both p<0.01).

Conclusions: CGM overestimated glycemic responses in numerous contexts. At times this can mischaracterize the GI. In addition, there is inter-individual heterogeneity of the accuracy of CGM to estimate fasting glucose concentrations. Correction for this difference reduces, but does not eliminate, postprandial overestimate of glycemia by CGM. Caution should be applied when inferring absolute or relative glycemic responses to foods using CGM, and capillary sampling should be prioritised for accurate quantification of glycemic response.

Clinical Trial Registry: https://clinicaltrials.gov/study/NCT06333184
Original languageEnglish
JournalAmerican Journal of Clinical Nutrition
DOIs
Publication statusPublished - 26 Feb 2025

Data Availability Statement

Data described in the manuscript, code book, and analytic code will be made publicly and freely available without restriction at https://doi.org/10.17632/3m82byt9s3.1

Fingerprint

Dive into the research topics of 'Continuous glucose monitor overestimates glycemia, with the magnitude of bias varying by postprandial test and individual – A randomized crossover trial.'. Together they form a unique fingerprint.

Cite this