PURPOSE: This study tests the hypothesis that the design changes incorporated in the newer generation Triathlon posterior-stabilised TKA design result in kinematics that more closely reproduce the kinematics observed in healthy knees than those achieved by the older generation Scorpio posterior-stabilised TKA design. METHODS: Eleven patients with Triathlon posterior-stabilised TKA, twelve patients with Scorpio posterior-stabilised TKA, and 22 subjects with normal asymptomatic knees underwent fluoroscopic assessment of the knee during a step-up exercise and a weight-bearing deep knee bend. Two-dimensional and three-dimensional knee kinematics were assessed including the maximum flexion, the patella tendon angle (PTA), the patella flexion angle (PFA), the minimum distance between cam and post, and the tibio-femoral contact positions. RESULTS: The average maximum flexion achieved was 114° (SD 3°), 91° (SD 10°), and 143° (SD 14°) for the Triathlon, Scorpio, and Normal groups. The average cam/post mechanism engagement was at 63° (SD 24°) and 82° (SD 16°) for the Triathlon and Scorpio groups. The condylar contact points showed a paradoxical anterior slide for the Scorpio group which was not present in the Triathlon group. The PTA and PFA values of both implants showed significant differences from normal. CONCLUSION: Overall, the Triathlon implant design, as compared to Scorpio TKA, produced kinematics closer to that of normal knees as proposed by the hypothesis. However, despite being closer to normal, the kinematics exhibited by the Triathlon group were still different from normal. A comparison of kinematic performance, taking into account altered design parameters, will contribute to improved understanding and future design considerations. LEVEL OF EVIDENCE: Case-control study, retrospective, comparative study, Level III.
Original languageEnglish
Pages (from-to)2800-2809
JournalKnee Surgery, Sports Traumatology, Arthroscopy
Volume21
Issue number12
Early online date11 Oct 2012
DOIs
Publication statusPublished - Dec 2013

Fingerprint

Biomechanical Phenomena
Knee
Patellar Ligament
Patella
Weight-Bearing
Thigh
scorpio
Case-Control Studies
Retrospective Studies
Exercise

Cite this

Constraints in posterior-stabilised TKA kinematics : a comparison of two generations of an implant. / Pandit, Hemant; van Duren, Bernard Hendrik; Price, M.; Tilley, S.; Gill, Harinderjit Singh; Thomas, Neil P.; Murray, David W.

In: Knee Surgery, Sports Traumatology, Arthroscopy, Vol. 21, No. 12, 12.2013, p. 2800-2809.

Research output: Contribution to journalArticle

Pandit, Hemant ; van Duren, Bernard Hendrik ; Price, M. ; Tilley, S. ; Gill, Harinderjit Singh ; Thomas, Neil P. ; Murray, David W. / Constraints in posterior-stabilised TKA kinematics : a comparison of two generations of an implant. In: Knee Surgery, Sports Traumatology, Arthroscopy. 2013 ; Vol. 21, No. 12. pp. 2800-2809.
@article{d291d4cccecb41d9a24d209528acf2d1,
title = "Constraints in posterior-stabilised TKA kinematics: a comparison of two generations of an implant",
abstract = "PURPOSE: This study tests the hypothesis that the design changes incorporated in the newer generation Triathlon posterior-stabilised TKA design result in kinematics that more closely reproduce the kinematics observed in healthy knees than those achieved by the older generation Scorpio posterior-stabilised TKA design. METHODS: Eleven patients with Triathlon posterior-stabilised TKA, twelve patients with Scorpio posterior-stabilised TKA, and 22 subjects with normal asymptomatic knees underwent fluoroscopic assessment of the knee during a step-up exercise and a weight-bearing deep knee bend. Two-dimensional and three-dimensional knee kinematics were assessed including the maximum flexion, the patella tendon angle (PTA), the patella flexion angle (PFA), the minimum distance between cam and post, and the tibio-femoral contact positions. RESULTS: The average maximum flexion achieved was 114° (SD 3°), 91° (SD 10°), and 143° (SD 14°) for the Triathlon, Scorpio, and Normal groups. The average cam/post mechanism engagement was at 63° (SD 24°) and 82° (SD 16°) for the Triathlon and Scorpio groups. The condylar contact points showed a paradoxical anterior slide for the Scorpio group which was not present in the Triathlon group. The PTA and PFA values of both implants showed significant differences from normal. CONCLUSION: Overall, the Triathlon implant design, as compared to Scorpio TKA, produced kinematics closer to that of normal knees as proposed by the hypothesis. However, despite being closer to normal, the kinematics exhibited by the Triathlon group were still different from normal. A comparison of kinematic performance, taking into account altered design parameters, will contribute to improved understanding and future design considerations. LEVEL OF EVIDENCE: Case-control study, retrospective, comparative study, Level III.",
author = "Hemant Pandit and {van Duren}, {Bernard Hendrik} and M. Price and S. Tilley and Gill, {Harinderjit Singh} and Thomas, {Neil P.} and Murray, {David W.}",
year = "2013",
month = "12",
doi = "10.1007/s00167-012-2233-x",
language = "English",
volume = "21",
pages = "2800--2809",
journal = "Knee Surgery, Sports Traumatology, Arthroscopy",
issn = "0942-2056",
publisher = "Springer Verlag",
number = "12",

}

TY - JOUR

T1 - Constraints in posterior-stabilised TKA kinematics

T2 - a comparison of two generations of an implant

AU - Pandit, Hemant

AU - van Duren, Bernard Hendrik

AU - Price, M.

AU - Tilley, S.

AU - Gill, Harinderjit Singh

AU - Thomas, Neil P.

AU - Murray, David W.

PY - 2013/12

Y1 - 2013/12

N2 - PURPOSE: This study tests the hypothesis that the design changes incorporated in the newer generation Triathlon posterior-stabilised TKA design result in kinematics that more closely reproduce the kinematics observed in healthy knees than those achieved by the older generation Scorpio posterior-stabilised TKA design. METHODS: Eleven patients with Triathlon posterior-stabilised TKA, twelve patients with Scorpio posterior-stabilised TKA, and 22 subjects with normal asymptomatic knees underwent fluoroscopic assessment of the knee during a step-up exercise and a weight-bearing deep knee bend. Two-dimensional and three-dimensional knee kinematics were assessed including the maximum flexion, the patella tendon angle (PTA), the patella flexion angle (PFA), the minimum distance between cam and post, and the tibio-femoral contact positions. RESULTS: The average maximum flexion achieved was 114° (SD 3°), 91° (SD 10°), and 143° (SD 14°) for the Triathlon, Scorpio, and Normal groups. The average cam/post mechanism engagement was at 63° (SD 24°) and 82° (SD 16°) for the Triathlon and Scorpio groups. The condylar contact points showed a paradoxical anterior slide for the Scorpio group which was not present in the Triathlon group. The PTA and PFA values of both implants showed significant differences from normal. CONCLUSION: Overall, the Triathlon implant design, as compared to Scorpio TKA, produced kinematics closer to that of normal knees as proposed by the hypothesis. However, despite being closer to normal, the kinematics exhibited by the Triathlon group were still different from normal. A comparison of kinematic performance, taking into account altered design parameters, will contribute to improved understanding and future design considerations. LEVEL OF EVIDENCE: Case-control study, retrospective, comparative study, Level III.

AB - PURPOSE: This study tests the hypothesis that the design changes incorporated in the newer generation Triathlon posterior-stabilised TKA design result in kinematics that more closely reproduce the kinematics observed in healthy knees than those achieved by the older generation Scorpio posterior-stabilised TKA design. METHODS: Eleven patients with Triathlon posterior-stabilised TKA, twelve patients with Scorpio posterior-stabilised TKA, and 22 subjects with normal asymptomatic knees underwent fluoroscopic assessment of the knee during a step-up exercise and a weight-bearing deep knee bend. Two-dimensional and three-dimensional knee kinematics were assessed including the maximum flexion, the patella tendon angle (PTA), the patella flexion angle (PFA), the minimum distance between cam and post, and the tibio-femoral contact positions. RESULTS: The average maximum flexion achieved was 114° (SD 3°), 91° (SD 10°), and 143° (SD 14°) for the Triathlon, Scorpio, and Normal groups. The average cam/post mechanism engagement was at 63° (SD 24°) and 82° (SD 16°) for the Triathlon and Scorpio groups. The condylar contact points showed a paradoxical anterior slide for the Scorpio group which was not present in the Triathlon group. The PTA and PFA values of both implants showed significant differences from normal. CONCLUSION: Overall, the Triathlon implant design, as compared to Scorpio TKA, produced kinematics closer to that of normal knees as proposed by the hypothesis. However, despite being closer to normal, the kinematics exhibited by the Triathlon group were still different from normal. A comparison of kinematic performance, taking into account altered design parameters, will contribute to improved understanding and future design considerations. LEVEL OF EVIDENCE: Case-control study, retrospective, comparative study, Level III.

UR - http://www.scopus.com/inward/record.url?scp=84888301570&partnerID=8YFLogxK

UR - http://dx.doi.org/10.1007/s00167-012-2233-x

U2 - 10.1007/s00167-012-2233-x

DO - 10.1007/s00167-012-2233-x

M3 - Article

C2 - 23052124

VL - 21

SP - 2800

EP - 2809

JO - Knee Surgery, Sports Traumatology, Arthroscopy

JF - Knee Surgery, Sports Traumatology, Arthroscopy

SN - 0942-2056

IS - 12

ER -