Abstract
This research aims to explore the feasibility of capacitance-digital converter and impedance converter for measurement module in electrical capacitance tomography (ECT) system. ECT sensor used was a cylindrical sensor having 8 electrodes. Absolute capacitance measurement system based on Sigma Delta Capacitance-to-Digital-Converter AD7746 has been shown to produce measurement with high resolution. Whereas, capacitance measurement with wide range of frequency is possible using Impedance Converter AD5933. Comparison of measurement accuracy by both AD7746 and AD5933 with reference of LCR meter was evaluated. Biological matters represented in water and oil were treated as object reconstructed into image using linear back projection (LBP) algorithm.
Original language | English |
---|---|
Title of host publication | AIP Conference Proceedings |
Publisher | AIP Publishing |
Volume | 1817 |
Edition | 1 |
ISBN (Electronic) | 9780735414853 |
DOIs | |
Publication status | Published - 21 Feb 2017 |
Event | 1st International Symposium of Biomedical Engineering: Biomedical Engineering's Recent Progress in Biomaterials, Drugs Development, and Medical Devices, ISBE 2016 - Depok City, Indonesia Duration: 31 May 2016 → 1 Jun 2016 |
Conference
Conference | 1st International Symposium of Biomedical Engineering: Biomedical Engineering's Recent Progress in Biomaterials, Drugs Development, and Medical Devices, ISBE 2016 |
---|---|
Country/Territory | Indonesia |
City | Depok City |
Period | 31/05/16 → 1/06/16 |
ASJC Scopus subject areas
- General Physics and Astronomy