An in-vitro investigation into the cement pressurization achieved during insertion of four different femoral stems

C Gozzard, S Gheduzzi, A W Miles, I D Learmonth

Research output: Contribution to journalArticlepeer-review

6 Citations (SciVal)

Abstract

Adequate cement pressurization during stem insertion improves the interdigitation of cement into bone. This increases the strength of the cement-bone interface, thus contributing to the reduction of the incidence of aseptic loosening, the commonest cause of revision surgery. This in-vitro study compared the cement pressurization achieved during insertion of four different stems of equivalent sizes: the Elite Plus (DePuy, UK), C-Stern (DePuy, UK), Exeter (Stryker, USA), and CPS-Plus (Plus Orthopedics, Switzerland). The maximum pressures attained at the time of stem insertion were recorded at proximal, mid and distal stem levels. The Elite Plus generated significantly higher distal pressures than the other stems. The CPS-Plus generated significantly greater proximal cement pressures than the Elite Plus, C-Stem, and Exeter prostheses. The triple taper of the C-Stern increased the cement pressurization medial to the stem. The stem shape and the presence or absence of a proximal stem centralizer affect cement pressurization. The presence of a proximal stem centralizer, a large stem volume, and a lateralmedial taper are all factors associated with increased cement pressurization during stem insertion
Original languageEnglish
Pages (from-to)407-413
Number of pages7
JournalProceedings of the Institution of Mechanical Engineers, Part H - Journal of Engineering in Medicine
Volume219
Issue numberH6
Publication statusPublished - 2005

Fingerprint

Dive into the research topics of 'An in-vitro investigation into the cement pressurization achieved during insertion of four different femoral stems'. Together they form a unique fingerprint.

Cite this