Abstract
A theoretical investigation of different electroneurogram recording techniques using electrode cuffs is presented. A new screened tripole arrangement is proposed with a higher inherent signal to interference ratio than the true tripole, which also allows the nulling of the residual electromyogram signal. The reduction in interference is small because the electrode impedance is large compared to the source resistance.
Original language | English |
---|---|
Pages (from-to) | 1281-1284 |
Journal | IEEE Transactions on biomedical engineering |
Volume | 47 |
Issue number | 9 |
DOIs | |
Publication status | Published - Sept 2000 |