A convex lateral tibial plateau for knee replacement

J. V. Bare, H. S. Gill, D. J. Beard, D. W. Murray

Research output: Contribution to journalArticle

39 Citations (Scopus)

Abstract

Unicompartmental knee replacements have not performed as well in the lateral compartment as in the medial. This may be because the tibial components have flat or slightly concave surfaces which match the medial plateau but not the convex lateral plateau. The aim of this study was to find the optimal radius for a convex lateral tibial component.

Twelve normal lateral tibial plateau were retrieved at knee replacement, and their surface contour in their mid sagittal plane was determined. The optimal circle was fitted and its radius measured. A series of different shaped tibial components were superimposed. From published information about the position of the femoral condyle relative to the tibia in different degrees of flexion, the flexion gap at these angles was determined.

The average radius of the lateral tibial plateau was 40 mm. However, as the surface was polyradial it was not clear if this average radius would be optimal.

In full flexion, a flat tibial plateau distracted the knee by 8 mm (p < 0.001). A 75 mm radius spherical tibial plateau did not alter the knee kinematics significantly and gave rise to a change in joint distraction of 1.5 mm. Spherical tibial plateau of 50 mm and 25 mm radii significantly altered knee kinematics (p < 0.001) and resulted in changes in distraction of 3 mm and 4 mm respectively.

The optimal shape for a unicompartmental lateral tibial plateau is likely to be a spherical dome with radius of about 75 mm. The incorporation of this shape in the lateral side of a total knee replacement might improve its flexion.
Original languageEnglish
Pages (from-to)122-126
Number of pages5
JournalThe Knee
Volume13
DOIs
Publication statusPublished - 2006

Fingerprint

Knee
Knee Replacement Arthroplasties
Biomechanical Phenomena
Thigh
Tibia
Joints
Bone and Bones

Cite this

A convex lateral tibial plateau for knee replacement. / Bare, J. V.; Gill, H. S.; Beard, D. J.; Murray, D. W.

In: The Knee, Vol. 13, 2006, p. 122-126.

Research output: Contribution to journalArticle

Bare, J. V. ; Gill, H. S. ; Beard, D. J. ; Murray, D. W. / A convex lateral tibial plateau for knee replacement. In: The Knee. 2006 ; Vol. 13. pp. 122-126.
@article{ea5b93bfe62a4504839cb908db6dd9e3,
title = "A convex lateral tibial plateau for knee replacement",
abstract = "Unicompartmental knee replacements have not performed as well in the lateral compartment as in the medial. This may be because the tibial components have flat or slightly concave surfaces which match the medial plateau but not the convex lateral plateau. The aim of this study was to find the optimal radius for a convex lateral tibial component.Twelve normal lateral tibial plateau were retrieved at knee replacement, and their surface contour in their mid sagittal plane was determined. The optimal circle was fitted and its radius measured. A series of different shaped tibial components were superimposed. From published information about the position of the femoral condyle relative to the tibia in different degrees of flexion, the flexion gap at these angles was determined.The average radius of the lateral tibial plateau was 40 mm. However, as the surface was polyradial it was not clear if this average radius would be optimal.In full flexion, a flat tibial plateau distracted the knee by 8 mm (p < 0.001). A 75 mm radius spherical tibial plateau did not alter the knee kinematics significantly and gave rise to a change in joint distraction of 1.5 mm. Spherical tibial plateau of 50 mm and 25 mm radii significantly altered knee kinematics (p < 0.001) and resulted in changes in distraction of 3 mm and 4 mm respectively.The optimal shape for a unicompartmental lateral tibial plateau is likely to be a spherical dome with radius of about 75 mm. The incorporation of this shape in the lateral side of a total knee replacement might improve its flexion.",
author = "Bare, {J. V.} and Gill, {H. S.} and Beard, {D. J.} and Murray, {D. W.}",
year = "2006",
doi = "10.1016/j.knee.2005.09.001",
language = "English",
volume = "13",
pages = "122--126",
journal = "The Knee",
issn = "0968-0160",
publisher = "Elsevier",

}

TY - JOUR

T1 - A convex lateral tibial plateau for knee replacement

AU - Bare, J. V.

AU - Gill, H. S.

AU - Beard, D. J.

AU - Murray, D. W.

PY - 2006

Y1 - 2006

N2 - Unicompartmental knee replacements have not performed as well in the lateral compartment as in the medial. This may be because the tibial components have flat or slightly concave surfaces which match the medial plateau but not the convex lateral plateau. The aim of this study was to find the optimal radius for a convex lateral tibial component.Twelve normal lateral tibial plateau were retrieved at knee replacement, and their surface contour in their mid sagittal plane was determined. The optimal circle was fitted and its radius measured. A series of different shaped tibial components were superimposed. From published information about the position of the femoral condyle relative to the tibia in different degrees of flexion, the flexion gap at these angles was determined.The average radius of the lateral tibial plateau was 40 mm. However, as the surface was polyradial it was not clear if this average radius would be optimal.In full flexion, a flat tibial plateau distracted the knee by 8 mm (p < 0.001). A 75 mm radius spherical tibial plateau did not alter the knee kinematics significantly and gave rise to a change in joint distraction of 1.5 mm. Spherical tibial plateau of 50 mm and 25 mm radii significantly altered knee kinematics (p < 0.001) and resulted in changes in distraction of 3 mm and 4 mm respectively.The optimal shape for a unicompartmental lateral tibial plateau is likely to be a spherical dome with radius of about 75 mm. The incorporation of this shape in the lateral side of a total knee replacement might improve its flexion.

AB - Unicompartmental knee replacements have not performed as well in the lateral compartment as in the medial. This may be because the tibial components have flat or slightly concave surfaces which match the medial plateau but not the convex lateral plateau. The aim of this study was to find the optimal radius for a convex lateral tibial component.Twelve normal lateral tibial plateau were retrieved at knee replacement, and their surface contour in their mid sagittal plane was determined. The optimal circle was fitted and its radius measured. A series of different shaped tibial components were superimposed. From published information about the position of the femoral condyle relative to the tibia in different degrees of flexion, the flexion gap at these angles was determined.The average radius of the lateral tibial plateau was 40 mm. However, as the surface was polyradial it was not clear if this average radius would be optimal.In full flexion, a flat tibial plateau distracted the knee by 8 mm (p < 0.001). A 75 mm radius spherical tibial plateau did not alter the knee kinematics significantly and gave rise to a change in joint distraction of 1.5 mm. Spherical tibial plateau of 50 mm and 25 mm radii significantly altered knee kinematics (p < 0.001) and resulted in changes in distraction of 3 mm and 4 mm respectively.The optimal shape for a unicompartmental lateral tibial plateau is likely to be a spherical dome with radius of about 75 mm. The incorporation of this shape in the lateral side of a total knee replacement might improve its flexion.

UR - http://www.ncbi.nlm.nih.gov/entrez/query.fcgi?cmd=Retrieve&db=PubMed&dopt=Citation&list_uids=16403637

UR - http://dx.doi.org/10.1016/j.knee.2005.09.001

U2 - 10.1016/j.knee.2005.09.001

DO - 10.1016/j.knee.2005.09.001

M3 - Article

VL - 13

SP - 122

EP - 126

JO - The Knee

JF - The Knee

SN - 0968-0160

ER -